RESTORING THE FACTORY DEFAULT SETUPS
The many features in this section allow the user to customize the balance to
suit a particular application. However, in doing this it is possible to inadvert-
ently set up the balance in such a way that it does not operate as expected.
To reset the factory defaults so that the: MODE key will select display re-
sponse rate; PRINT key will print a stable reading; 2400 baud, no parity; all
units enabled with “x” units set to grams per second; perform the following
steps.
STEP 1.
TARE
POWER
MODE
FUnC
0.00
Hold down the TARE key then press the MODE key. Release both keys, the
display will read "FUnC".
STEP 2.
MODE
TARE
POWER
FACdEFFUnC
Press the TARE key repeatedly until the display reads "FACdEF" for factory
defaults.
Note: To escape anytime during this procedure, press the TARE key until
"ESCAPE" is displayed and press the MODE key.
STEP 3.
0.00FACdEF
MODE
TARE
POWER
Press the MODE key to restore the original factory defaults. The balance will
display “buSY” and then return to the normal weighing mode.
Note: Restoring the factory defaults will return your balance to all of the
factory span and temperature calibration settings. You MUST
recalibrate (span) your balance after restoring the factory defaults. If
you are experiencing a temperature induced offset, you should also
run the temperature compensation calibration procedure.